R&S®TS9982 EMS Test System Family
Radiated and conducted EMS measurements in line with commercial, wireless, automotive and MIL standards
System configuration
The
The EMS control unit contains a signal generator, a path selector, a power meter and an interlock circuit. The use of the R&S®IMS, which combines all these features in one unit, makes for an extremely compact design. Due to the few system components, a rack is usually not necessary.
For large systems with special requirements, such as those involved when complete vehicles are measured, single components are used for the EMS control unit. In this case, the core of the system is the R&S®TS-RSP switching unit, which is configured as necessary by means of different plug-ins and the possibility to drive external power switches up to 10 kW.
Based on experience gained from the large number of systems implemented worldwide, all system components have been ideally matched to one another, yielding efficient field generation on the DUT.
In the 1 GHz to 40 GHz frequency range, a mobile microwave test system is used directly adjacent to the DUT, thus preventing high RF cable loss above 3 GHz.
An interlock circuit that switches off the RF power when the door of the hall is opened, for example, provides personal safety and is always included in the system.