6241A DC Voltage/Current Source/Monitor

(NOT distributed in North America & Asia.)

Specifications

6241A

Voltage Source/Measurement Range
Range Source Range Setting Resolution Measurement Range Measurement Resolution
300 mV 0 to ±320.00 mV 10 µV 0 to ±320.999 mV 1 µV
3 V 0 to ±3.2000 V 100 µV 0 to±3.20999 V 10 µV
30 V 0 to ±32.000 V 1 mV 0 to±32.0999 V 100 µV
Line regulation: ±0.003% of range or less
Load regulation: ±0.003% of range or less (at 4-wire connection with maximum load)
Current Source/Measurement Range
Range Source Range Setting Resolution Measurement Range Measurement Resolution
30 µA 0 to ±32.000 µA 1 nA 0 to ±32.0999 µA 100 pA
300 µA 0 to ±320.00 µA 10 nA 0 to ±320.999 µA 1 nA
3 mA 0 to ±3.2000 mA 100 nA 0 to ±3.20999 mA 10 nA
30 mA 0 to ±32.000 mA 1 µA 0 to ±32.0999 mA 100 nA
300 mA 0 to ±320.00 mA 10 µA 0 to ±320.999 mA 1 µA
500 mA 0 to ±500.00 mA 20 µA 0 to ±500.999 mA 1 µA
Resistance Measurement Range
  Measurement Range Measurement Resolution
Determined by the operation of Voltage range / Current range 0 Ω to 1.6 GΩ Minimum 2 µΩ
Output Noise
For voltage source within the range from no load to maximum load [Vp-p]
Voltage Source:
Range Load resistance Low Frequency Noise High Frequency Noise
    DC to 100Hz DC to 10kHz DC to 20MHz
300 mV - 50µV 200µV 3mV
3 V - 50µV 300µV 3mV
30V/6V - 500µV 2mV 4mV
General Source/Measurement Functions
DC Source/Measurement Source/measurement of DC voltage/current
Pulse Source/Measurement Source/measurement of pulse voltage/current (However, measurement auto range in pulse source is impossible.)
DC Sweep Source/Measurement Source/measurement with linear, two-slope linear, random and fixed level.
Pulse Sweep Source/Measurement Source/measurement with linear, two-slope linear, random and fixed level.
(However, measurement auto range in pulse source is impossible.)
Integration Time 100µs, 500µs, 1ms, 5ms, 10ms, 1PLC, 100ms,
200ms, or S/H
S/H: Measuring in Sample & Hold Mode (Integration time: 100µs), (Enabled only in pulse source and pulse sweep source modes.)
PLC: Power Line Cycle:
50Hz:20ms 60Hz:16.66ms
Sweep Mode Reverse ON/OFF (forward to backward / one way)
Sweep Repeat Time 1 to 1000 times, or infinite
Number of Sweep Maximum Step 8000 steps
Random Sweep Maximum Memory 8000 data
Measurement Data Memory 8000 data
Measurement Auto Range Available only in VSIM or ISVM
Measurement Function Link Mode Links the source function to the measurement function (VSIM or ISVM) ON/OFF available
Limiter Possible to set separately at HI or LO sides (However, in case of current limiter, the same polarity can not be set.)
Calculation Functions NULL calculation
Comparator calculation (HI/GO/LO)
Scaling calculation
MAX/MIN/AVE/TOTAL calculation
Trigger Style Auto-Trigger, external Trigger
Output Terminal Front; Safety sockets,
HI OUTPUT, HI SENSE, LO OUTPUT, LO SENSE
Maximum Input Voltage between Terminals R6241A: 32V peak MAX (between HI and LO)
2V peak MAX /(between OUTPUT and SENSE)
250V MAX (between LO and chassis)
Maximum Remote Sensing Voltage ± 1V MAX
between HI OUTPUT and HI SENSE
between LO OUTPUT and LO SENSE
(The voltage between HI SENSE and LO SENSE must be within the maximum output voltage range)
GPIB Interface Compliant with IEEE-488.2-1987
Interface functions: SH1, AH1,T5, L4, SR1, RL1, PP0, DC1, DT1, C0, E2
Connector: Amphenol 24 pin
USB Interface USB 2.0 Full-Speed; Connector Typ B
External Control Signal TRIGGER IN,
INTERLOCK, OPERATE IN/OUT, COMPLETE OUT, SYNC OUT
General Specification
Power Supply: AC power 100V, 120V, 220V, 240C (switchable by user)
Line Frequency: 50Hz/60Hz
Power Consumption: 6241A: 95VA or less
Dimensions: Approx. 212(W) x 88(H) x 400(D) mm
Mass: 6241A: 6kg or less
Safety: IEC 61010-1
EMI: EN61326 class A